Health product recall

Xper Flex Cardio Physiomonitoring System-Main Unit

Starting date:
August 21, 2017
Posting date:
September 11, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Hospitals, Healthcare Professionals
Identification number:
RA-64496

Reason

Because of intermittent communication between the host system and the flex cardio, it is possible for any of the following conditions to occur:

  • boom monitor (display) may not display all active waveform and/or vital sign data.
  • delayed audible and visual alarms at the flex cardio and boom monitor respectively, due to a mismatch between the alarm limits on the host system and the flex cardio/boom monitor.
  • after patient admission and case initiation at the host system, the case may not be fully transmitted to the flex cardio. In this case, the boom monitor in the exam/procedure room will display vital sign data, but the flex cardio will not provide audible alarms.
  • after patient discharge and case termination at the host system, case termination may not be transmitted to the flex cardio. In this case, the boom monitor in the exam/procedure room will continue to display vital sign data, and the flex cardio will continue to provide audible alarms.

Affected products

Xper Flex Cardio Physiomonitoring System-Main Unit

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

  • FC2010
  • FC2020

Companies

Manufacturer
Invivo, a Div. of Philips Medical Systems
12151 Research Parkway
Orlando
32826
Florida
UNITED STATES