This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

XN-10 SERIES HEMATOLOGY ANALYZER (2016-06-24)

Starting date:
June 24, 2016
Posting date:
July 7, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-59256

Affected products

A. XN-10 SERIES HEMATOLOGY ANALYZER

Reason

Sysmex corporation, Japan (SCJ) discovered that some XN-Series analyzers' Reaction Chamber Heaters have a possibility of abnormal heating that can lead to wire disconnection, and a scorched or burnt odour.

Affected products

A. XN-10 SERIES HEMATOLOGY ANALYZER

Lot or serial number

serial numbers 17371 - 17784

Model or catalog number

XN-10

Companies
Manufacturer

Sysmex Corporation

1-5-1 Wakinohama-kaigandori

Chuo-ku

JAPAN