This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

XN-10 MODULE (2016-06-23)

Starting date:
June 23, 2016
Posting date:
July 7, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Hospitals, Healthcare Professionals
Identification number:
RA-59254

Affected products

A. XN-10 MODULE

Reason

Sysmex Corporation, Japan (SCJ) determined that some XN-Series analyzers had the wiring for the feeder assembly #113 improperly routed, potentially causing contact with moving parts of the tube sampling mechanism. Contact with the piercer up/down mechanism can cause abrasion of the wire's coating and a short circuit of the tube sampling mechanism.

Affected products

A. XN-10 MODULE

Lot or serial number

More than 10 numbers, contact manufacturer.  

Model or catalog number

 XN-10

Companies
Manufacturer

Sysmex Corporation

1-5-1 Wakinohama-kaigandori

Chuo-ku

JAPAN