Health product recall

XIA 3 System - Serrato - Polyaxial Screw (2019-06-18)

Starting date:
June 18, 2019
Posting date:
July 5, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-70365


Last updated:

Affected Produtcs

XIA 3 System - Serrato - Polyaxial Screw

Reason

Stryker identified a serrato screw of discrepant length during inspection. Subsequently a product manufacturing mix up/swap between lot # B90243 [9.5X100mm] and lot # B88751 [9.5X90mm]. Lot B88751 90mm screws were manufactured with a 100mm length instead of the required 90mm length; lot B90243 100mm screws were manufactured with a 90mm length instead of the required 100mm length.

Affected products

XIA 3 System - Serrato - Polyaxial Screw

Lot or serial number
  • B88751
  • B90243
Model or catalog number

4826195100

Companies
Manufacturer

Stryker Spine

2 PEARL COURT

ALLENDALE

07401

New Jersey