Health product recall

Xhibit Central Station (2019-09-09)

Starting date:
September 9, 2019
Posting date:
October 3, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71123



Last updated: 2019-10-03

Affected Products

Xhibit Central Station

Reason

Spacelabs has received reports of losing alarm audio after a power failure or cable disconnection and reconnection. Visual alarms and alarm printouts perform normally.

Affected products

Xhibit Central Station

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

96102

Companies
Manufacturer

Spacelabs Healthcare Inc.

35301 SE Center Street

Snoqualmie

98065

Washington

UNITED STATES