X-Guide X-Mark Probe Tool
Brand(s)
Last updated
Summary
Product
X-Guide X-Mark Probe Tool
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
X-Guide X-Mark Probe Tool |
2007010010 |
P010673 |
Issue
The lower weld of the tip of the X-Mark Probe Tool may be missing. If the upper weld breaks during use, it will release the tip of the probe tool, potentially causing a patient to swallow or aspirate.
Recall start date: October 7, 2022
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Dental
Companies
X-Nav Technologies, LLC.
1555 Bustard Road, Lansdale, Pennsylvania, United States, 19446
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-72050