Health product recall

WS80A Diagnostic Ultrasound System, HS70A Ultrasonic Diagnostic Apparatus (2019-06-24)

Starting date:
June 24, 2019
Posting date:
July 5, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-70375

Last updated: 2019-07-05

Affected Products

  1. WS80A Diagnostic Ultrasound System
  2. HS70A Ultrasonic Diagnostic Apparatus

Reason

A specific mode of operation of certain versions of WS80A/HS70A systems may pose a risk to patient safety. On June 13, 2019, the manufacturer discovered that probes may overheat when decreasing the Doppler SV (Sample Volume) size value set by the user in the Doppler Only mode on WS80A/HS70A systems. There has been no actual injury reported to Samsung Medison.

Affected products

A. WS80A Diagnostic Ultrasound System

Lot or serial number

S1JUM3HM300001K

Model or catalog number

WS80A

Companies
Manufacturer
SAMSUNG MEDISON CO., LTD.
3366 HANSEO-RO,, NAM-MYEON
HONGCHEON-GUN
25108
SOUTH KOREA

B. HS70A Ultrasonic Diagnostic Apparatus

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

HS70A

Companies
Manufacturer
SAMSUNG MEDISON CO., LTD.
3366 HANSEO-RO,, NAM-MYEON
HONGCHEON-GUN
25108
SOUTH KOREA