WS80A Diagnostic Ultrasound System, HS70A Ultrasonic Diagnostic Apparatus (2019-06-24)
- Starting date:
- June 24, 2019
- Posting date:
- July 5, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-70375
Last updated: 2019-07-05
Affected Products
- WS80A Diagnostic Ultrasound System
- HS70A Ultrasonic Diagnostic Apparatus
Reason
A specific mode of operation of certain versions of WS80A/HS70A systems may pose a risk to patient safety. On June 13, 2019, the manufacturer discovered that probes may overheat when decreasing the Doppler SV (Sample Volume) size value set by the user in the Doppler Only mode on WS80A/HS70A systems. There has been no actual injury reported to Samsung Medison.
Affected products
A. WS80A Diagnostic Ultrasound System
Lot or serial number
S1JUM3HM300001K
Model or catalog number
WS80A
Companies
- Manufacturer
-
SAMSUNG MEDISON CO., LTD.
3366 HANSEO-RO,, NAM-MYEON
HONGCHEON-GUN
25108
SOUTH KOREA
B. HS70A Ultrasonic Diagnostic Apparatus
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
HS70A
Companies
- Manufacturer
-
SAMSUNG MEDISON CO., LTD.
3366 HANSEO-RO,, NAM-MYEON
HONGCHEON-GUN
25108
SOUTH KOREA