Winnipeg Ventilator 2.0 (2021-04-06)



Starting date:
April 6, 2021
Posting date:
April 29, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-75487


Last updated:

Affected Products 

Winnipeg Ventilator 2.0

Reason

In ongoing product reliability testing it was identified that the linear sensor output, under certain rare conditions, may cause the device to stop ventilating, display error code 1605 and sound an alarm indicating a technical fault.

Affected products

Winnipeg Ventilator 2.0

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

204427

Companies
Manufacturer

CANADIAN EMERGENCY VENTILATORS INC

455 BOLESKINE ROAD

VICTORIA

V8Z 1E7

British Columbia

CANADA