Health product recall

Winnipeg Ventilator 2.0 (2021-04-06)

Starting date:
April 6, 2021
Posting date:
April 29, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-75485



Last updated:
2021-04-29

Affected Products 

Winnipeg Ventilator 2.0

Reason

A post-production review identified that a limited number of devices from early production have two bolts that were torqued to a previous lower design specification

Affected products

Winnipeg Ventilator 2.0

Lot or serial number

More than 10 numbers, contact manufacturer. 

Model or catalog number

204427

Companies
Manufacturer

CANADIAN EMERGENCY VENTILATORS INC

455 BOLESKINE ROAD

VICTORIA

V8Z 1E7

British Columbia

CANADA