WASHER/DISINFECTOR (2020-12-22)
- Starting date:
- December 22, 2020
- Posting date:
- January 7, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- Healthcare Professionals, General Public, Hospitals
- Identification number:
- RA-74705
Last updated:
2021-01-07
Affected Products
WASHER/DISINFECTOR
Reason
Getinge has discovered a water inlet plastic hose connection that may deform or break due to material fatigue or hose clamp overtightening. A broken hose connection can cause a water leakage and water spray.
Affected products
WASHER/DISINFECTOR
Lot or serial number
31262
31272
31273
31274
31305
W50027773
W50030134
W50036817
WAA077309
Model or catalog number
CM320
Companies
- Manufacturer
-
Getinge Disinfection AB
Ljungadalsgatan 11
Växjö
35246
SWEDEN