Health product recall

Wall Mounted Charging Station (2021-04-27)



Starting date:
April 27, 2021
Posting date:
June 10, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-75783


Last updated:

Affected Products 

Wall Mounted Charging Station

Reason

Air Liquide Medical Systems has been informed of an isolated incident in Germany regarding Monnal T60 used on a wall charging station. An excessive overheating has appeared in their neighborhood (the plastic housing around the contact points having been altered and plastic housing found melted).

Affected products

Wall Mounted Charging Station

Lot or serial number

All lots. 

Model or catalog number

KA010300

Companies
Manufacturer

Air Liquide Medical Systems

6 Rue Georges Besse, Parc de Haute Technologie

Antony

92182

FRANCE