Health product recall

WalkAway plus system - 96 - instrument (2019-02-28)

Starting date:
February 28, 2019
Posting date:
March 22, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-69370

Affected Products

WalkAway plus system - 96 - instrument

Reason

As a result of an internal investigation for electromagnetic compatibility (EMC) requirements testing, Beckman Coulter has confirmed that WalkAway plus system - 96 - instrument can reset and lose communication when delivered with a 2 kV power surge.

Affected products

WalkAway plus system - 96 - instrument

Lot or serial number

39630290

Model or catalog number

B1018-284

Companies
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES