WalkAway plus system - 96 - instrument (2019-02-28)
- Starting date:
- February 28, 2019
- Posting date:
- March 22, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-69370
Affected Products
WalkAway plus system - 96 - instrument
Reason
As a result of an internal investigation for electromagnetic compatibility (EMC) requirements testing, Beckman Coulter has confirmed that WalkAway plus system - 96 - instrument can reset and lose communication when delivered with a 2 kV power surge.
Affected products
WalkAway plus system - 96 - instrument
Lot or serial number
39630290
Model or catalog number
B1018-284