Health product recall

VNS Therapy SenTiva® Duo Generator

Last updated

Summary

Product
VNS Therapy SenTiva® Duo Generator
Issue
Medical devices - Performance
What to do

Customers to be notified. Instructions to follow on response. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

VNS Therapy SenTiva® Duo Generator

More than 10 numbers, contact manufacturer.

1000-D

Issue

Complaints have been received on the SenTiva Generator reporting a loss of therapy or inability to make therapy changes with the VNS therapy programming system. A reed switch malfunction, resulting in the reed switch sticking in a closed position has been identified as the cause of these device events.

Recall start date: January 10, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Neurology
Companies
Livanova Usa, Inc.
100 Cyberonics Blvd., Houston, Texas, United States, 77058
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76908

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe