Health product recall

Vivo 45 LS

Brand(s)
Last updated

Summary

Product
Vivo 45 LS
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Vivo 45 LS

S/N: 1R070KF to 1R100F9

230000

Issue

In devices with the affected firmware versions, treatment in volume-controlled ventilation modes (VCV, VCV-SIMV and VCV-MPV) can, under specific trigger conditions, result in delivered tidal volume being reduced significantly below the set value.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies

Breas Medical AB

Foretagsvagen 1, Molnlycke, Sweden, 43533

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82020