Health product recall

VITROS® XT Chemistry Product ALB-TP Slides

Last updated

Summary

Product
VITROS® XT Chemistry Product ALB-TP Slides
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

VITROS® XT Chemistry Product ALB-TP Slides

1900148359
1900179815
1700120620

684 4293

Issue

QuidelOrtho™ has confirmed an issue involving VITROS® XT Chemistry Products ALB-TP slides which may create dust and debris within the microslide subsystem on VITROS® XT 3400 Chemistry Systems and VITROS® XT 7600 integrated systems. Dust and debris from XT ALB-TP slides may impact VITROS® Chemistry Products Na+ slides leading to a potential increase in non-reproducible, positively, or negatively biased Na+ results.

Recall start date: July 12, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies

Ortho-Clinical Diagnostics Inc.

100 Indigo Creek Drive, Rochester, New York, United States, 14626

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75871

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe