Health product recall

Vitros Xt Systems

Last updated

Summary

Product
Vitros Xt Systems
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
Vitros Xt 3400 Chemistry System VERSION 3.7.2 AND BELOW 684 4458
Vitros Xt 7600 Integrated System VERSION 3.7.2 AND BELOW 6844461

Issue

Ortho Clinical Diagnostics (Ortho) became aware of an anomaly that affects Vitros Xt 3400 and Vitros Xt 7600 systems. These systems utilize a digital imaging reflectometer and led light source to generate Vitros microslide assay results. Ortho's on-going investigation has determined that performing a system shutdown on Vitros Xt 3400 and Vitros Xt 7600 systems may, in some instances, cause a positive or negative shift in the electrical current supplied to the led used in conjunction with the digital imaging reflectometer. This shift in electrical current may also lead to a shift in multiple Vitros microslide assay results reported by the system.

Recall start date: July 22, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies

Ortho-Clinical Diagnostics Inc.

100 Indigo Creek Drive, Rochester, New York, United States, 14626

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-64425