Vitros Xt Systems
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
---|---|---|
Vitros Xt 3400 Chemistry System |
VERSION 3.7.2 AND BELOW |
684 4458 |
Vitros Xt 7600 Integrated System |
VERSION 3.7.2 AND BELOW |
6844461 |
Issue
Ortho Clinical Diagnostics (Ortho) became aware of an anomaly that affects Vitros Xt 3400 and Vitros Xt 7600 systems. These systems utilize a digital imaging reflectometer and led light source to generate Vitros microslide assay results. Ortho's on-going investigation has determined that performing a system shutdown on Vitros Xt 3400 and Vitros Xt 7600 systems may, in some instances, cause a positive or negative shift in the electrical current supplied to the led used in conjunction with the digital imaging reflectometer. This shift in electrical current may also lead to a shift in multiple Vitros microslide assay results reported by the system.
Recall start date: July 22, 2022