VITROS XT 7600 Integrated System (2019-03-20)
- Starting date:
- March 20, 2019
- Posting date:
- April 5, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-69500
Last updated: 2019-04-05
Affected Products
VITROS XT 7600 Integrated System
Reason
Due to an undetected mechanical interference, a MicroSlide may not be properly positioned into the Tip Locator. If this occurs, the sample may not be properly dispensed on the slide. Ortho Clinical Diagnostics identified during internal testing that the threshold used in locating the dispense blade position in the Tip Locator is incorrect. If an undetected MicroSlide position error occurs, a condition code is inadvertently not reported by system and the potentially erroneous test result is generated. VITROS XT MicroSlides, MicroWell and MicroTip assays are NOT affected.
Affected products
VITROS XT 7600 Integrated System
Lot or serial number
Software Versions 3.4 and 3.4.1.
Model or catalog number
- 684 4461
Companies
- Manufacturer
-
Ortho-Clinical Diagnostics Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES