Health product recall

VITROS ECi Immunodiagnostic Products (2018-11-01)

Starting date:
November 1, 2018
Posting date:
November 28, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68370

Affected products

  1. VITROS ECi Immunodiagnostic Products NTproBNP Reagent Pack
  2. VITROS ECi Immunodiagnostic Products TSH Reagent Pack
  3. VITROS ECi Immunodiagnostic Products Prolactin Reagent Pack
  4. VITROS ECi Immunodiagnostic Products FSH Reagent Pack
  5. VITROS ECi Immunodiagnostic Products Folate Reagent Controls & Calibrators
  6. VITROS ECi Immunodiagnostic Products Free PSA Reagent Pack

Reason

Ortho Clinical Diagnostics determined that biased results may occur for specific VITROS Immunodiagnostic Products (MicroWell Assays) at biotin concentrations which are lower than indicated in the current Instructions for Use (IFU).

Historically, these assays were evaluated for the interference consistent with CLSI document EP7. Biotin was tested and was not found to cause a bias of >/=10%.

To assess the potential interference of biotin supplements on VITROS MicroWell Assays, Ortho performed more robust testing using the new CLSI guidelines EP7 and EP37 and determined that the information in the current IFUs is no longer supported for these five (5) assays. The concentration of biotin that causes >/=10% bias is listed in the customer letter.

Affected products

A. VITROS ECi Immunodiagnostic Products NTproBNP Reagent Pack

Lot or serial number

ALL EXPIRED/IN-DATE/FUTURE LOT

Model or catalog number

680 2156

Companies
Manufacturer

ORTHO-CLINICAL DIAGNOSTICS

Felindre Meadows, Pencoed, Bridgend

Great Britain

CF35 5PZ

UNITED KINGDOM



B. VITROS ECi Immunodiagnostic Products TSH Reagent Pack

Lot or serial number

ALL EXPIRED/IN-DATE/FUTURE LOT

Model or catalog number

191 2997

Companies
Manufacturer

Ortho Clinical Diagnostics

Felindre Meadows

Pencoed, Bridgend

CF35 5PZ

UNITED KINGDOM



C. VITROS ECi Immunodiagnostic Products Prolactin Reagent Pack

Lot or serial number

ALL EXPIRED/IN-DATE/FUTURE LOT

Model or catalog number

184 9793

Companies
Manufacturer

Ortho Clinical Diagnostics

Felindre Meadows

Pencoed, Bridgend

CF35 5PZ

UNITED KINGDOM



D. VITROS ECi Immunodiagnostic Products FSH Reagent Pack

Lot or serial number

ALL EXPIRED/IN-DATE/FUTURE LOT

Model or catalog number

193 1922

Companies
Manufacturer

Ortho Clinical Diagnostics

Felindre Meadows

Pencoed, Bridgend

CF35 5PZ

UNITED KINGDOM



E. VITROS ECi Immunodiagnostic Products Folate Reagent Controls & Calibrators

Lot or serial number

ALL EXPIRED/IN-DATE/FUTURE LOT

Model or catalog number

151 3266

Companies
Manufacturer

Ortho Clinical Diagnostics

Felindre Meadows

Pencoed, Bridgend

CF35 5PZ

UNITED KINGDOM



F. VITROS ECi Immunodiagnostic Products Free PSA Reagent Pack

Lot or serial number

ALL EXPIRED/IN-DATE/FUTURE LOT

Model or catalog number

684 2845

Companies
Manufacturer

Ortho Clinical Diagnostics

Felindre Meadows

Pencoed, Bridgend

CF35 5PZ

UNITED KINGDOM