Health product recall

Vitros Chemistry Products System - Crp Slides (C-Reactive Protein)

Last updated

Summary

Product
Vitros Chemistry Products System - Crp Slides (C-Reactive Protein)
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Vitros Chemistry Products System - Crp Slides (C-Reactive Protein)

More than 10 numbers, contact manufacturer.

192 6740

Issue

QuidelOrtho™ received complaints on the affected coating of VITROS® CRP slides, related to lower-than-expected results when testing patient samples known to have elevated concentrations of CRP (>90 mg/l or >9 mg/dl). When tested neat, results were reported as being within the upper end of the measuring range (70-90 mg/l or 7-9 mg/dl) however, upon dilution, results for these samples were reported as having concentrations greater than 2x the neat result.

Patient samples with a CRP concentration above 90 mg/l (9 mg/dl) (typically above 150 mg/l (15 mg/dl)) may be reported as falsely reduced when samples are tested neat, with the dilute sample results predicting greater than 2x greater than the neat result. The observed bias will vary; however, the predicted neat CRP results will generally range between 70-90 mg/l (7-9 mg/dl).

Recall start date: October 25, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Immunology
Companies
Ortho-Clinical Diagnostics Inc.
100 Indigo Creek Drive, Rochester, New York, United States, 14626
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76455