VITROS Chemistry Products 5,1 FS Chemistry System - dLDL Assay (2018-10-08)
- Starting date:
- October 8, 2018
- Posting date:
- November 2, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-68184
Affected products
VITROS Chemistry Products 5,1 FS Chemistry System - dLDL Assay
Reason
Ortho Clinical Diagnostics (Ortho) has initiated this recall following complaints received for GENs 26 and 27 as well as internal testing of a new lot that determined that GENs 26 and 27 are negatively biased by approximately 15 to 20 mg/dL (0.38 to 0.51 mmol/L) across the measuring range.
Affected products
VITROS Chemistry Products 5,1 FS Chemistry System - dLDL Assay
Lot or serial number
- GEN: 26-6396
- GEN: 26-6639
- GEN: 27-6808
Model or catalog number
680 1728
Companies
- Manufacturer
-
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
UNITED STATES