Health product recall

VITROS 5, 1 FS Chemistry System, VITROS 3600 Immunodiagnostic System, VITROS 5600 Integrated System, and VITROS 4600 Chemistry System (2018-10-21)

Starting date:
October 21, 2018
Posting date:
November 15, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68222

Affected products

  1. VITROS 5,1 FS Chemistry System
  2. VITROS 5,1 FS Chemistry System (Refurbished)
  3. VITROS 3600 Immunodiagnostic System
  4. VITROS 5600 Integrated System
  5. VITROS 4600 Chemistry System

Reason

Under specific scenarios, a software anomaly will occur causing the VITROS Systems to incorrectly determine the reagent's on-analyzer stability time. When the anomaly occurs, the system could use reagents past their on-analyzer stability time without alerting the user with a condition code or flagging the associated results with an RE (Reagent Expired) code. In the event reagents are used beyond their specific on-analyzer stability times, the associated test results may be affected.

Affected products

  1. VITROS 5,1 FS Chemistry System

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

680 1375

Companies
Manufacturer
Ortho Clinical Diagnostics Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES

B. VITROS 5,1 FS Chemistry System (Refurbished)

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

6801890

Companies
Manufacturer
Ortho Clinical Diagnostics Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES

C. VITROS 3600 Immunodiagnostic System

Lot or serial number

J36000773
J36000942
J36001003
J36001126

Model or catalog number

680 2783
680 2914

Companies
Manufacturer
Ortho Clinical Diagnostics Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES

D. VITROS 5600 Integrated System

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

680 2413
680 2915

Companies
Manufacturer
Ortho Clinical Diagnostics Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES

E. VITROS 4600 Chemistry System

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

680 2445

Companies
Manufacturer
Ortho Clinical Diagnostics Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES