VITEK® 2 Systems
Brand(s)
Summary
BioMérieux recommends confirmatory testing of resistant results for Enterobacterales/Enterobacteriaceae and P. aeruginosa on VITEK® 2 GN-AST cards for select antibiotics.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
VITEK 2 System |
Not applicable. |
AST-N804 |
VITEK 2 System |
Not applicable. |
AST-XN42 |
VITEK 2 Compact 30 Systems |
Not applicable. |
Not applicable. |
VITEK 2 Compact 60 System |
Not applicable. |
Not applicable. |
Issue
BioMérieux identified increased out-of-range high quality control results for the following antibiotic formulations: ceftazidime/avibactam (CZA02n), imipenem (IPM05n), imipenem/relebactam (IPR01n), meropenem/vaborbactam (MEV01n), and polymyxin B (PB02n), as well as reports of false resistance on VITEK® 2 GN-AST Cards when testing Enterobacterales/Enterobacteriaceae and P. aeruginosa.
Additional information
Details
| BioMerieux, Inc. |
| 100 Rodolphe Street, Durham, North Carolina, United States, 27712 |