This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Vitek 2 System - Susceptibility Test Cards
Brand(s)
Last updated
Summary
Product
Vitek 2 System - Susceptibility Test Cards
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Vitek 2 System - Susceptibility Test Cards | 1402761504 1412754504 | 423340 423341 |
Issue
The St. Louis manufacturing site discovered a calculation error in the master formulation for ceftriaxone (cro02n) used in specific card lots. The error was related to the ceftriaxone concentration used in two wells of the test card.
Recall start date: March 15, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Microbiology
Companies
| Biomerieux, Inc. |
| 100 Rodolphe Street, Durham, North Carolina, United States, 27712 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75312
Get notified
Receive emails about new and updated recall and safety alerts.