This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Vitek 2 System - Susceptibility Test Cards
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Vitek 2 System - Susceptibility Test Cards |
1402761504 |
423340 |
Issue
The St. Louis manufacturing site discovered a calculation error in the master formulation for ceftriaxone (cro02n) used in specific card lots. The error was related to the ceftriaxone concentration used in two wells of the test card.
Recall start date: March 15, 2024
Additional information
Details
| Biomerieux, Inc. |
| 100 Rodolphe Street, Durham, North Carolina, United States, 27712 |
Get notified