Vitek 2 System - Susceptibility Test Cards
Brand(s)
Last updated
Summary
Product
Vitek 2 System - Susceptibility Test Cards
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
---|---|---|
Vitek 2 System - Susceptibility Test Cards |
1402761504 |
423340 |
Issue
The St. Louis manufacturing site discovered a calculation error in the master formulation for ceftriaxone (cro02n) used in specific card lots. The error was related to the ceftriaxone concentration used in two wells of the test card.
Recall start date: March 15, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Microbiology
Companies
Biomerieux, Inc. |
100 Rodolphe Street, Durham, North Carolina, United States, 27712 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75312
Get notified