This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

VITEK 2 AST-GN27 Card

Starting date:
March 1, 2012
Posting date:
February 4, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-20293

Recalled Products

A. VITEK 2 AST-GN27 Card

Reason

Biomerieux received reports from some customers about a high rate card or antibiotic terminations for VITEK 2 AST-GN27 card (#22245). An internal investigation revealed sporadic scratches on the outer surface of the gray transfer tubing of some card of the lot 267215010.

Affected products

A. VITEK 2 AST-GN27 Card

Lot or serial number

267215010

Model or catalog number

22245

Companies
Manufacturer

Biomerieux, Inc.