Health product recall

VIDAS LYME IGG II (2016-03-14)

Starting date:
March 14, 2016
Posting date:
April 4, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-57772

Affected Products

VIDAS LYME IGG II

Reason

Some customers reported that they could not perform calibration. Results of investigation reveal that calibration cannot be performed due to S1 RFV signal which triggers "invalid calibration" error.

Affected products

VIDAS LYME IGG II

Lot or serial number
  • 1004127190
Model or catalog number
  • 417401
Companies
Manufacturer

Biomerieux Sa

Chemin De L'orme

Marcy-l'etoile

69280