Health product recall

VERTICAL DRAIN TUBE ATTACHMENT DEVICE (2021-02-11)



Starting date:
February 11, 2021
Posting date:
February 26, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-75051


Last updated:

Affected Products 

VERTICAL DRAIN TUBE ATTACHMENT DEVICE

Reason

The vertical drain tube attachment device is designed to stabilize and secure a variety of catheters, surgical drains, and tubes from sizes 5-40 French. The tube holder can separate from the barrier that secures the device to the patients' body. This separation might contribute to the potential risk of the tube/drain/catheter migration or loss.

Affected products

VERTICAL DRAIN TUBE ATTACHMENT DEVICE

Lot or serial number

0K02

1A09

1A10

Model or catalog number

9782

Companies
Manufacturer

HOLLISTER INCORPORATED

2000 HOLLISTER DRIVE

LIBERTYVILLE

60048

Illinois

UNITED STATES