Summary
Product
Versacross Connect LAAC Access Solution
Issue
Medical devices - Performance issue
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
---|---|---|
Versacross Connect LAAC Access Solution |
More than 10 numbers, contact manufacturer. |
VXAK0007 |
Issue
Boston Scientific Corporation is initiating a product advisory to inform customers of the increased potential for J-tipped mechanical guidewires packaged from 10 March to 28 July 2023 within the Baylis Medical Company Versacross Connect LAAC Access Solution kit, to get stuck in Versacross Connect dilators when advancing or retracting the mechanical guidewire.
Recall start date: September 26, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies
Baylis Medical Company Inc. |
5959 Trans-Canada Highway, Montreal, Quebec, Canada, H4T 1A1 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74427