Health product recall

Vaser Amplifier

Starting date:
June 15, 2017
Posting date:
July 7, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-63860

Reason

Solta Medical is conducting a voluntary recall of the Vaser Amplifier that were returned to the Solta Bothell Washington service depot for repair between September 8, 2016 and May 23, 2017 due to the potential that the device was inaccurately tuned by the service depot during repair. The use of the product in the affected range constitutes a potential risk and could contribute to burns.

Affected products

Vaser Amplifier

Lot or serial number

SN 000V9Z

SN 0019XJ

Model or catalog number

110-0032

Companies

Manufacturer

Solta Medical Inc.