This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

VARISOURCE IX AFTERLOADER (2015-09-29)

Starting date:
September 29, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-56046

Affected Products

VARISOURCE IX AFTERLOADER

Reason

The manufacturer has been made aware of an application freeze occurring during treatment initiation. When a forced logoff occurs while the dummy wire is extended during a treatment session, the partial fraction generated by the system will be incorrect.

Affected products

VARISOURCE IX AFTERLOADER

Lot or serial number
  • Version 1.2 Build 2.4336
  • Version 1.2.1. Build 3726
Model or catalog number

AL20090000

Companies
Manufacturer

Varian Medical Systems, Inc.

3100 Hansen Way

Palo Alto

94304

California

UNITED STATES