VARISOURCE IX AFTERLOADER (2015-09-29)
- Starting date:
- September 29, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-56046
Affected Products
VARISOURCE IX AFTERLOADER
Reason
The manufacturer has been made aware of an application freeze occurring during treatment initiation. When a forced logoff occurs while the dummy wire is extended during a treatment session, the partial fraction generated by the system will be incorrect.
Affected products
VARISOURCE IX AFTERLOADER
Lot or serial number
- Version 1.2 Build 2.4336
- Version 1.2.1. Build 3726
Model or catalog number
AL20090000