VariSoft
Brand(s)
Last updated
Summary
Product
VariSoft
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Industry
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
VariSoft |
5388368 |
1002827 |
Issue
Unomedical A/S has found that in rare cases the VariSoft infusion set connector detaches more easily from the infusion set than expected, requiring less force to disconnect than intended, thereby interrupting the delivery of insulin.
Recall start date: October 11, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Unomedical A/S
Aaholmvej 1-3, Osted, Lejre, Denmark, 4320
Published by
Health Canada
Audience
Industry
Recall class
Type II
Identification number
RA-74511