Health product recall

Variax 2 Locking Screw - Non-Sterile

Brand(s)
Last updated

Summary

Product
Variax 2 Locking Screw - Non-Sterile
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Variax 2 Locking Screw - Non-Sterile

AM5253

657328

Issue

Stryker has confirmed that one lot of Variax® 2 Locking crews was manufactured without the hexalobular feature required for mating with the associated screwdriver.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Orthopaedics
Companies

Stryker Gmbh

Bohnackerweg 1, Selzach, Solothurn, Switzerland, 2545

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-81913

Get notified

Receive emails about new and updated recall and safety alerts.