Health product recall

Vanta/Sequentia Lt Clinician Programmer Application

Brand(s)
Last updated

Summary

Product
Vanta/Sequentia Lt Clinician Programmer Application
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Vanta/Sequentia Lt Clinician Programmer Application

NMS503505H
NMS502986H
NMS503341H
NMS503484H
NMS503162H
NMS502821H
NMS502807H
NMS503187H
NMS503195H
NMS502816H
NMS502805H
NMS502845H
NMS502840H
NMS502849H
NMS502814H
NMS502820H
NMS502356H

A71200

Issue

In rare occasions, the Vanta patient programmer app (PP APP) a72200 (all versions) can display a "system update needed, service code 303" during the initial interrogation of a Vanta implantable neurostimulator (INS) model 977006. When this code is observed, therapy cannot be adjusted and/or turned off with the PP APP.

Recall start date: November 29, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies

Medtronic Inc.

8200 Coral Sea Street N.E., Mounds View, Minnesota, United States, 55112

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-72396

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe