Health product recall

Vanta/Sequentia Lt Clinician Programmer Application

Brand(s)
Last updated

Summary

Product
Vanta/Sequentia Lt Clinician Programmer Application
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Vanta/Sequentia Lt Clinician Programmer Application

More than 10 numbers, contact manufacturer.

A71200

Issue

Connection issue for Vanta pain pc implantable neurostimulators when transplanted in the same patient. "Too many devices" error will appear from Clinician Programmer Application (CPA) when there is more than 1 Vanta ins detected by the clinician telemetry module. Current Vanta labeling does not include information with respect to what the "too many devices found" informational message means or how to troubleshoot it.

Recall start date: August 4, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Medtronic Inc.

8200 Coral Sea Street N.E., Mounds View, Minnesota, United States, 55112

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-64472

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe