Vanta/Sequentia Lt Clinician Programmer Application
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
---|---|---|
Vanta/Sequentia Lt Clinician Programmer Application |
More than 10 numbers, contact manufacturer. |
A71200 |
Issue
Connection issue for Vanta pain pc implantable neurostimulators when transplanted in the same patient. "Too many devices" error will appear from Clinician Programmer Application (CPA) when there is more than 1 Vanta ins detected by the clinician telemetry module. Current Vanta labeling does not include information with respect to what the "too many devices found" informational message means or how to troubleshoot it.
Recall start date: August 4, 2022