Health product recall

Vantage Titan HSR (2018-05-09)

Starting date:
May 9, 2018
Posting date:
June 1, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66944

Affected products

Vantage Titan HSR

Reason

When an operator uses the respiratory gating with steady mode on in spin echo sequence scan, the SAR value displayed on the console may be lower than the actual SAR value.

Affected products

Vantage Titan HSR

Lot or serial number

Y5A10Z2001

Model or catalog number

MRT-1504_/Y5

Companies
Manufacturer
Toshiba Medical Systems Corporation
1385 Shimoishigami
Otawara-Shi, Tochigi-ken
324-8550
JAPAN