Product
Vantage Orian-Main Unit
Issue
Medical devices - Labelling and packaging
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Vantage Orian-Main Unit

All lots.

MRT-1550/YX
MRT-1550/XD
MRT-1550/YP

Issue

It has been found that some of the maximum SFG (spatial field gradient) values displayed in the operation window and some of the maximum SFG values described in the safety manual are different from the designed values.

Recall start date: December 12, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Canon Medical Systems Corporation

1385, Shimoishigami, Otawara-Shi, Tochigi-Ken, Japan, 324-8550

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74836