Vantage Orian
Brand(s)
Last updated
Summary
Product
Vantage Orian
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Vantage Orian |
YXB2182001 |
MRT-1550/YP |
Issue
It has been determined that the terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed. Because of this, the cable became disconnected, and arcing occurred. Melting of the cable due to this arcing caused the cable securing parts below it to ignite, resulting in fire at the terminal block.
Recall start date: February 16, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
| Canon Medical Systems Corporation |
| 1385, Shimoishigami, Otawara-Shi, Tochigi-Ken, Japan, 324-8550 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75185
Get notified