Vantage Galan 3T
Brand(s)
Last updated
Summary
Product
Vantage Galan 3T
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Vantage Galan 3T |
Y5C1922001 |
MRT-3020/Y5 |
Issue
It has been found that if there is slow leakage of refrigerant gas from the superconducting magnet over a long period of time, air ingress into the helium vessel via the leak point may result in ice forming from moisture in the air. In the worst case, if the ice is not eliminated, the helium tank could rupture, resulting in a helium leak.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
Canon Medical Systems Corporation
1385, Shimoishigami, Otawara-Shi, Tochigi-Ken, Japan, 324-8550
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-78458
Get notified