Vantage Galan 3T
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Vantage Galan 3T |
Y5C1922001 |
MRT-3020/Y5 |
Issue
It has been determined that the terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed. Because of this, the cable became disconnected, and arcing occurred. Melting of the cable due to this arcing caused the cable securing parts below it to ignite, resulting in fire at the terminal block.
Recall Start Date: November 14, 2023
Additional information
Details
Canon Medical Systems Corporation
1385, Shimoishigami, Otawara-Shi, Tochigi-Ken, Japan, 324-8550
Get notified