Health product recall

Vanadis Core T21/T18/T13 Reagent Cartridge

Brand(s)
Last updated

Summary

Product
Vanadis Core T21/T18/T13 Reagent Cartridge
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Vanadis Core T21/T18/T13 Reagent Cartridge

746100
745859
745619
744592
742254
745232
746522
741096
744481
744811
746077
743926
745136
744491

3222-001B_CA

Issue

Affected lots can include plates, which are slightly higher than normally, and the plates may rest higher on the instrument adapter (plates manufactured by our supplier using a different mold). This may result in run failure due to issues with moving the plate within the instrument or, due to suboptimal temperature during extraction of cell-free DNA, can cause insufficient protein degradation, which can potentially affect performance of the Vanadis Core® T21/T18/T13 reagent cartridge 3222-001B_CA.

Recall Start Date: August 3, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies

Wallac Oy

Mustionkatu 6, Turku, Finland, 20750

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74118

Get notified

Receive emails about new and updated recall and safety alerts.