Valleylab FT10 FT Series Energy Platform
Brand(s)
Last updated
Summary
Product
Valleylab FT10 FT Series Energy Platform
Issue
Medical devices - Device compatibility
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalog number |
|---|---|---|
| Valleylab FT10 FT Series Energy Platform | More than 10 numbers, contact manufacturer. | VLFT10GEN |
Issue
Upon insertion of a new (unused) Ligasure™ device, the Valleylab™ FT10 energy platform running software versions 4.0.1, 4.0.2 and 4.0.3 may erroneously indicate that the Ligasure™ device was used previously. When this occurs, the energy platform will display error "e420 usage limit" or error "e416 unknown instrument," and the Ligasure™ device would not be allowed to be used.
Recall Start Date: November 20, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies
Covidien Llc
15 Hampshire Street, Mansfield, Massachusetts, United States, 02048
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74682