Health product recall

Valleylab FT10 FT Series Energy Platform

Brand(s)
Last updated

Summary

Product
Valleylab FT10 FT Series Energy Platform
Issue
Medical devices - Device compatibility
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Valleylab FT10 FT Series Energy Platform

More than 10 numbers, contact manufacturer.

VLFT10GEN

Issue

Upon insertion of a new (unused) Ligasure™ device, the Valleylab™ FT10 energy platform running software versions 4.0.1, 4.0.2 and 4.0.3 may erroneously indicate that the Ligasure™ device was used previously. When this occurs, the energy platform will display error "e420 usage limit" or error "e416 unknown instrument," and the Ligasure™ device would not be allowed to be used.

Recall Start Date: November 20, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies

Covidien Llc

15 Hampshire Street, Mansfield, Massachusetts, United States, 02048

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74682

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe