Vacutainer Brand Eclipse Blood Collection Needle (2019-08-02)
- Starting date:
- August 2, 2019
- Posting date:
- August 16, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-70741
Last updated: 2019-08-16
Affected Products
Vacutainer Brand Eclipse Blood Collection Needle
Reason
Becton Dickinson and Company has initiated a medical device recall of specific production lots of BD Vacutainer Eclipse Blood Collection Needle 22G x 1.25". Becton Dickinson and Company has indicated that lots 9025826, 8354527 may have the bevel missing from the non-patient needle end of the Eclipse Blood Collection Needle. This non-patient needle is housed in a sleeve that encloses the needle after puncture to prevent any leakage. If there is no bevel, the needle may damage the sleeve, causing leakage.
Affected products
Vacutainer Brand Eclipse Blood Collection Needle
Lot or serial number
8354527
9025826
Model or catalog number
368608
Companies
- Manufacturer
-
Becton Dickinson and Company
1 Becton Drive
Franklin Lakes
07417
New Jersey
UNITED STATES