V60 VENTILATOR, SYMBOLS ONLY, WITH CFLEX AND AVAPS OPTIONS Serviced With Touchscreen RP Kit 453561511951 (2019-09-18)
- Starting date:
- September 18, 2019
- Posting date:
- October 25, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-71367
Last updated: 2019-10-25
Affected Products
V60 VENTILATOR, SYMBOLS ONLY, WITH CFLEX AND AVAPS OPTIONS Serviced With Touchscreen RP Kit 453561511951
Reason
If a patient requires a change in therapy, there is no warning associated with an unresponsive touchscreen, and therefore the clinician would have no prior indication of a touchscreen failure and would be unable to change ventilator settings.
Affected products
V60 VENTILATOR, SYMBOLS ONLY, WITH CFLEX AND AVAPS OPTIONS Serviced With Touchscreen RP Kit 453561511951
Lot or serial number
Not applicable.
Model or catalog number
1053614
Companies
- Manufacturer
-
Respironics California Inc.
2271 Cosmos Court
Carlsbad
92011
California
UNITED STATES