V60 Ventilator, Symbols Only, with CFLEX and AVAPS Options (2019-09-18)
- Starting date:
- September 18, 2019
- Posting date:
- October 3, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-71145
Last updated: 2019-10-03
Affected Products
V60 Ventilator, Symbols Only, with CFLEX and AVAPS Options
Reason
The touchscreen must be used to change certain therapy settings. The touchscreens in affected units may become frozen and fail to respond to touch commands. Consequently, if a patient requires a change in a ventilator therapy setting, e.g., FIO2, the change could not be made. Although it would be immediately apparent to the user that the changes cannot be made, there is no advance warning of touchscreen failure.
Affected products
V60 Ventilator, Symbols Only, with CFLEX and AVAPS Options
Lot or serial number
Not applicable.
Model or catalog number
1053614
Companies
- Manufacturer
-
Respironics California LLC
2271 Cosmos court
Carlsbad
92011
California
UNITED STATES