This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

V Series Monitoring System (2014-01-27)

Starting date:
January 27, 2014
Posting date:
March 18, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-38309

Recalled Products

V Series Monitoring System 

Reason

The V dock power pins connecting the V series monitor to the V dock may lose spring tension, potentially causing the screen to temporarily go blank, lose communication with the central station (when one is in use) or shut down.

Affected products

V Series Monitoring System (2014-01-27)

Lot or serial number

EGL19008944

EGL19008961

Model or catalog number

0998-00-1801-01

Companies
Manufacturer

Mindray Ds Usa, Inc.

800 Macarthur Blvd.

Mahwah

07430

New Jersey

UNITED STATES