This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

User-Fillable Reagent Bottle (2015-05-06)

Starting date:
May 6, 2015
Posting date:
January 25, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-56772

Affected Devices

User-Fillable Reagent Bottle

Reason

Serial numbers were repeated on different bottles on this lot of User-Fillable Reagent Bottles. If Autostainer Link instrument reads a serial number that has been assigned previously, a bottling error will occur.

Affected products

User-Fillable Reagent Bottle

Lot or serial number

10092763

Model or catalog number

SK200

Companies
Manufacturer

Dako North America, Inc.

6392 via Real

Carpinteria

93013

California

UNITED STATES