This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
User-Fillable Reagent Bottle (2015-05-06)
- Starting date:
- May 6, 2015
- Posting date:
- January 25, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-56772
Affected Devices
User-Fillable Reagent Bottle
Reason
Serial numbers were repeated on different bottles on this lot of User-Fillable Reagent Bottles. If Autostainer Link instrument reads a serial number that has been assigned previously, a bottling error will occur.
Affected products
User-Fillable Reagent Bottle
Lot or serial number
10092763
Model or catalog number
SK200
Companies
- Manufacturer
-
Dako North America, Inc.
6392 via Real
Carpinteria
93013
California
UNITED STATES