UniCel DxI 600 and 800 Access Immunoassay Systems
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
UniCel DxI 800 Access Immunoassay - Analyzer-Class 3 |
More than 10 numbers, contact manufacturer. |
A71457 |
UniCel DxI 800 Access Immunoassay - Analyzer-Class 3 |
More than 10 numbers, contact manufacturer. |
A71456 |
UniCel DxI 800 Access Immunoassay - Analyzer-Class 3 |
More than 10 numbers, contact manufacturer. |
973100 |
UniCel DxI 600 Access Immunoassay Sys - Instrument W/ Spot B Gantry Class 2 |
More than 10 numbers, contact manufacturer. |
A71460 |
UniCel DxI 600 Access Immunoassay Sys - Instrument W/ Spot B Gantry Class 2 |
More than 10 numbers, contact manufacturer. |
A71461 |
UniCel DxI 600 Access Immunoassay Sys - Instrument W/Spot B Gantry Class 3 |
More than 10 numbers, contact manufacturer. |
A71460 |
UniCel DxI 600 Access Immunoassay Sys - Instrument W/Spot B Gantry Class 3 |
More than 10 numbers, contact manufacturer. |
A71461 |
UniCel DxI 600 Access Immunoassay System - Instrument Class 3 |
900950 |
A30260 |
UniCel DxI 600 Access Immunoassay System - Instrument Class 3 |
902386 |
A30260 |
UniCel DxI 600 Access Immunoassay System - Instrument Class 3 |
900807 |
A30260 |
UniCel DxI 600 Access Immunoassay System - Instrument Class 3 |
901719 |
A30260 |
UniCel DxI 600 Access Immunoassay System - Instrument Class 3 |
902624 |
A30260 |
UniCel DxI 600 Access Immunoassay System - Instrument Class 3 |
900570 |
A30260 |
UniCel DxI 600 Access Immunoassay System - Instrument Class 3 |
902385 |
A30260 |
UniCel DxI 600 Access Immunoassay System - Instrument Class 3 |
902622 |
A30260 |
UniCel DxI 600 Access Immunoassay System - Instrument Class 3 |
901684 |
A30260 |
UniCel DxI 600 Access Immunoassay System - Instrument Class 3 |
902475 |
A30260 |
UniCel DxI 600 Access Immunoassay System |
900950 |
A30260 |
UniCel DxI 600 Access Immunoassay System |
902386 |
A30260 |
UniCel DxI 600 Access Immunoassay System |
900807 |
A30260 |
UniCel DxI 600 Access Immunoassay System |
901719 |
A30260 |
UniCel DxI 600 Access Immunoassay System |
902624 |
A30260 |
UniCel DxI 600 Access Immunoassay System |
902385 |
A30260 |
UniCel DxI 600 Access Immunoassay System |
900570 |
A30260 |
UniCel DxI 600 Access Immunoassay System |
902622 |
A30260 |
UniCel DxI 600 Access Immunoassay System |
901684 |
A30260 |
UniCel DxI 600 Access Immunoassay System |
902475 |
A30260 |
UniCel DxI 600 Access Immunoassay System - Instrument Class 2 |
900950 |
A30260 |
UniCel DxI 600 Access Immunoassay System - Instrument Class 2 |
902386 |
A30260 |
UniCel DxI 600 Access Immunoassay System - Instrument Class 2 |
900807 |
A30260 |
UniCel DxI 600 Access Immunoassay System - Instrument Class 2 |
901719 |
A30260 |
UniCel DxI 600 Access Immunoassay System - Instrument Class 2 |
902624 |
A30260 |
UniCel DxI 600 Access Immunoassay System - Instrument Class 2 |
900570 |
A30260 |
UniCel DxI 600 Access Immunoassay System - Instrument Class 2 |
902385 |
A30260 |
UniCel DxI 600 Access Immunoassay System - Instrument Class 2 |
902622 |
A30260 |
UniCel DxI 600 Access Immunoassay System - Instrument Class 2 |
901684 |
A30260 |
UniCel DxI 600 Access Immunoassay System - Instrument Class 2 |
902475 |
A30260 |
UniCel DxI 800 Access Immunoassay System - Analyzer - Class II |
More than 10 numbers, contact manufacturer. |
A71457 |
UniCel DxI 800 Access Immunoassay System - Analyzer - Class II |
More than 10 numbers, contact manufacturer. |
A71456 |
UniCel DxI 800 Access Immunoassay System - Analyzer - Class II |
More than 10 numbers, contact manufacturer. |
973100 |
Issue
Beckman Coulter UniCel DxI 600 and 800 instruments fail to start after rebooting because the AuObj.exe software does not launch. A specific LED pattern on MCC cards confirms a hardware startup fault. This non-conformance requires FSE intervention.
Additional information
Details
Beckman Coulter, Inc.
250 S. Kraemer Blvd., Brea, California, United States, 92821