Health product recall

Unicel Dxi 600 Access Immunoassay System- Thyroglobulin Assay

Last updated

Summary

Product
Unicel Dxi 600 Access Immunoassay System- Thyroglobulin Assay
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Mode or catalogue number

Unicel Dxi 600 Access Immunoassay System- Thyroglobulin Assay

439163

33860

Issue

Multiple customers reported receiving high patient results when using access thyroglobulin reagent pack lot 439163. An internal review of batch records determined this was the first lot released using antibodies produced by BEC instead of a third party supplier.

Recall start date: July 2, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Chemistry
Companies
Beckman Coulter, Inc.
250 S. Kraemer Blvd., Brea, California, United States, 92821
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77759

Get notified

Receive emails about new and updated recall and safety alerts.