Health product recall

UniCel DxI 600 Access Immunoassay System - Intrinsic Factor AB (2019-04-13)

Starting date:
April 13, 2019
Posting date:
May 3, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-69834


Last updated:

Affected Products

A. UniCel DxI 600 Access Immunoassay System - Intrinsic Factor AB

B. Access System - Intrinsic Factor AB

C. Intrinsic Factor AB Assay

Reason

Beckman Coulter has determined for the Access Intrinsic Factor Antibody (IFab) assay that in rare cases a "negative" result interpretation, rather than "equivocal", is assigned when the result is between 1.20 and 1.2000000476 AU/mL.

Affected products

A. UniCel DxI 600 Access Immunoassay System - Intrinsic Factor AB

Lot or serial number

All lots.

Model or catalog number

387992

Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

California

UNITED STATES



B. Access System - Intrinsic Factor AB

Lot or serial number

All lots.

Model or catalog number

387992

Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

California

UNITED STATES



C. Intrinsic Factor AB Assay

Lot or serial number

All lots.

Model or catalog number

387992

Companies
Manufacturer

Beckman Coulter Inc.

250 S. Kraemer Blvd.

Brea

92821

California

UNITED STATES