Issue:
Medical Devices
Audience:
General Public (GP), Healthcare Professionals, Hospitals
Identification number:
RA-15136

Recalled Products:

Unicel DXH Slidemaker Stainer Coulter Cellular Analysis System, Part Number 775222, Software Revision 1.5.1.0

Reason

The DXH SMS (Slidemaker Stainer) System is generating an "invalid code" message (error code 75) when the LIS transmits any field in the record that is not populated (nil). The LIS will retransmit the test order to the analyzer and the DXH SMS will reject the retransmission and send a message to the LIS that the test order was already received/processed.

Affected products

Unicel DXH Slidemaker Stainer Coulter Cellular Analysis System, Part Number 775222, Software Revision 1.5.1.0

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

775222

Companies
Manufacturer

Beckman Coulter Inc.