Health product recall

UNICEL DXH SLIDEMAKER STAINER (2016-08-01)

Starting date:
August 1, 2016
Posting date:
August 19, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-59884

Affected products

A. UNICEL DXH SLIDEMAKER STAINER

Reason

Beckman Coulter has received and confirmed a report of a fire within the Stainer module of a customer's DxH Slidemaker Stainer. The fire was fully contained within the stainer module. The event was preceded by a significant increase in the frequency of the automatic top-off of the module's methanol bath. Beckman Coulter is initiating this field action to reduce the potential for flammable liquids coming into contact with electrical components in the stainer module possibly causing a fire.

Affected products

A. UNICEL DXH SLIDEMAKER STAINER

Lot or serial number

All lots

Model or catalog number

775222